Performance Veri
fi
cation
Table 15: DC voltage accuracy with remote sense
Instrument test
voltage
DUT voltage
Test current
Min
Measured
Max
Channel 1
0%
0.00000 V
0.5 A
-0.01000 V
__________
0.01000 V
25%
7.5000 V
0.5 A
7.48775 V
__________
7.51225 V
50%
15.00000 V
0.5 A
14.98550 V
__________
15.01450 V
75%
0.00000 V
2.5 A
-0.01000 V
__________
0.01000 V
100%
30.0000 V
0.5 A
29.98100 V
__________
30.01900 V
Channel 2
0%
0.00000 V
0.5 A
-0.01000 V
__________
0.01000 V
25%
7.5000 V
0.5 A
7.48775 V
__________
7.51225 V
50%
15.00000 V
0.5 A
14.98550 V
__________
15.01450 V
75%
22.50000 V
0.5 A
22.48325 V
__________
22.51675 V
100%
30.00000 V
0.5 A
29.98100 V
__________
30.01900 V
Channel 3
0%
0.00000 V
2.5 A
-0.01000 V
__________
0.01000 V
25%
1.5000 V
0.5 A
14.98550 V
__________
15.01450 V
50%
3.00000 V
2.5 A
2.98910 V
__________
3.01090 V
75%
4.50000 V
2.5 A
4.48865 V
__________
4.51135 V
100%
6.00000 V
2.5 A
5.98820 V
__________
6.01180 V
16