Speci
fi
cations
Table 13: Safety characteristics
Item
Description
U.S. Nationally Recognized Testing Laboratory (NRTL) Listing
UL61010-1–2004 Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use
Canadian Certi
fi
cation
CAN/CSA C22.2 No. 61010-1-2004 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
European Union Compliance
Low Voltage Directive 2006/95/EC
EN61010-1 2001 Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use
Safety Certi
fi
cations
Additional Compliance
IEC61010-1 2001 Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use
Channel to Channel Isolation
Voltage
Any output can be
fl
oated up to 240 V (DC + peak AC with AC limited to a maximum 3V
pk-pk
and
maximum of 60 Hz) relative to any other channel terminal. Remote sense terminals should
remain within 1 V of their respective outputs to avoid possible damage.
Channel to Ground Isolation
Voltage
Any output can be
fl
oated up to 240 V (DC + peak AC with AC limited to a maximum 3V
pk-pk
and
maximum of 60 Hz) relative to the Earth Ground terminal. Remote sense terminals should
remain within 1 V of their respective outputs to avoid possible damage.
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
11