Appendix B: Kelvin (4W) S530 system diagnostics
1. If the current row is A and B:
Use pin 1 to connect the force and sense terminals of the current SMU under test to row A.
Use pin 2 to connect the force and sense terminals of the support SMU (except GNDU) to row B.
Use pin 3 to connect the force and sense terminals of the common lo (GNDU) to row C.
Use the current row (A or B) on the sense card to connect the current SMU under test and the
support SMU together.
Figure 45: Example SMU sense test connections using rows A and B
2. If the current row is C, D, E, and F:
Use pin 1 to connect the force and sense terminals of the current SMU to row C.
Use pin 2 to connect the force and sense terminals of the support SMU (except GNDU) to row D.
Use pin 3 to connect the force and sense terminals of the common lo (GNDU) to row E.
Use the current row (C, D, E, or F) on the sense card to connect the SMU under test and the test
support SMU together.
Figure 46: SMU sense test connections using rows C, D, and E