38
Model 7706 module
:CODE
(:CALibration:PROTected:CODE)
NOTES
The :CODE command should be sent only once before performing calibration. Do not send :CODE before
each calibration step.
The code parameter must be enclosed in single quotes.
Purpose
To program the calibration code or password so that you can perform the Model 7706
calibration procedures.
Format
:CAL:PROT:CODE '<CHAR_STRING>'
Parameter
Up to a 8-character string including letters and numbers.
Description
The :CODE command enables the Model 2700 calibration procedures when performing
these procedures over the bus. This command must be sent to the unit before sending any
other Model 7706 calibration command. The default calibration code is KI002700.
Example
:CAL:PROT:CODE 'KI002700'
Send default code of KI002700.
:COUNt?
(:CALibration:PROTected:CARD1:COUNt?)
(:CALibration:PROTected:CARD2:COUNt?)
Purpose
To determine how many times a Model 7706 has been calibrated.
Format
:CAL:PROT:CARD1:COUN?
:CAL:PROT:CARD2:COUN?
Response
<n> Calibration count.
Description
The :CARD1:COUNt? and :CARD2:COUNt? commands allow you to determine how
many times a Model 7706 in Slot 1 and Slot 2, respectively, has been calibrated.
NOTE
Use the :COUNt? command to help you monitor for unauthorized calibration procedures.
Example
:CAL:PROT:CARD1:COUN?
Request card 1 calibration count.