21
Connection
1 Special keys and power switch:
Normal operation.
2 Function and operation keys:
Top Row
Unshifted
DCV, ACV, DCI, ACI,
Ω
2,
Ω
4, FREQ, TEMP
Blocked.
Shifted
RATIO,CH-AVG, CONT, OCOMP, PERIOD
Blocked.
Middle Row
Unshifted
TRIG
Triggers the unit to read the current Totalizer count.
and
Moves between Totalizer channels (if more than one Model 7706
installed).
STORE, RECALL, FILTER, REL
Blocked.
Shifted
LIMITS
Sets the High Limit count for the Totalizer.
ON/OFF
Enables/disables limits.
MONITOR
Sets unit to monitor Totalizer channel and turns monitor on. If monitor is
already on, it will be turned off.
CARD
Enters the card specific menu.
DELAY, HOLD
Normal Operation.
TYPE, CH-OFF
Blocked.
Bottom Row
Unshifted
OPEN, CLOSE
Normal operation.
EXIT
Clears the Totalizer count to zero.
STEP, SCAN, DIGITS, RATE, ENTER
Blocked.
Shifted
SAVE, SETUP, CONFIG, TEST, GPIB, RS-232
Normal operation.
HALT, LSYNC
Blocked.
3 Range keys:
and
Normal operation.
AUTO Blocked.
4 INPUTs switch:
Normal operation—the position of this switch does not affect Totalizer operation.
Adding Totalizer to scan list (via front panel)
Toggle the state of the Totalizer channel in the scan list using the SHIFT->CH-OFF key. The TYPE and EDGE controls are set
globally through the SHIFT->CARD menu key.
NOTE
Changing Totalizer settings (“READ” or “RRES”) must be done outside of the scanlist configuration.
Adding to scan list via GPIB
The ROUT:SCAN command is used to add Totalizer to the scan list. When a card that supports Totalizer is in the unit, the To-
talizer channel can only be used as a Totalizer. Whenever it is added to the list it will read the current count value. The TYPE
and EDGE controls are set globally through the SENSe:TOTalize:EDGE and SENSe:TOTalize:TYPE commands.