![Keithley 7706 User Manual Download Page 44](http://html2.mh-extra.com/html/keithley/7706/7706_user-manual_3982827044.webp)
40
Model 7706 module
:LOCK?
(:CALibration:PROTected:CARD1:LOCK?)
Purpose
To read Model 7706 calibration lock status.
Format
:CAL:PROT:CARD1:LOCK?
Response
0 Calibration locked.
1 Calibration unlocked.
Description
The :LOCK? query requests status from the Model 2700 on Model 7706 calibration locked/
unlocked state. Calibration must be enabled sending the :CODE command before
calibration can be performed.
Example
:CAL:PROT:CARD1:LOCK?
Request card 1 cal lock state.
:SAVE
(:CALibration:PROTected:CARD1:SAVE)
Purpose
To save calibration constants in card EEROM after the calibration procedure.
Format
:CAL:PROT:CARD1:SAVE
Parameter
None
Description
The :SAVE command stores calculated calibration constants derived during Model 7706
calibration in card EEROM. (EEROM is non-volatile memory.) Calibration constants will
be retained indefinitely once saved.Generally, :SAVE is sent after all other calibration steps
(except for :LOCK).
NOTE
Card calibration will be only temporary unless the :SAVE command is sent to permanently store calibration
constants.
Example
:CAL:PROT:CARD1:SAVE
Save card calibration constants.