3-16
Step 3. Enter Scan List.
Switching Card
For a non-matrix switching card,
enter the scan list (Channels 1, 3, 4, 5, and 6) for slot 1
as follows:
Note: The following procedure assumes that slot 1 is
configured for non-matrix switching. If using slot 2 for
non-matrix switching, modify (A) of steps 1 and 2 by
pressing “2” instead of “1”.
1. Enter first channel (Channel 1) using the single
channel entry method:
A. Press “1”
Enters slot
B. Press “1”
Enters channel
C. Press “
” or “ENTER” Enters channel sepa-
rator (comma)
SCAN CHANNELS 1!1,
2. Enter the rest of the channels (Channels 3 through
6) using the channel range entry method:
A. Press “1”
Enters slot
B. Press “3”
Enters first channel
C. Press “-”
Enters range designator and
slot
D. Press “6”
Enters last channel
E. Press “
” or
Enters channel list terminator
“ENTER”
(comma)
SCAN CHANNELS 1!1, 1!3-1!6,
Note: Channels 3, 4, 5 and 6 could have been entered
using the single channel entry method demonstrated
in step 1.
Matrix Card
For a matrix card, enter the scan list
(Row 1 Column 1, Row 2 Column 2, Row 2 Column 3,
Row 2 Column 4, and Row 2 Column 5 for slot 2) as fol-
lows:
Note: The following procedure assumes that slot 2 is
configured for a matrix. If using slot 1 for a matrix,
modify (A) of steps 1 and 2 by pressing “1” instead of
“2”.
1. Enter first channel (Row 1 Column 1) using the sin-
gle channel entry method:
A. Press “2”
Enters slot
B. Press “1”
Enters row
C. Press “1”
Enters column
D. Press “
” or
Enters channel separa-
“ENTER”
tor (comma)
SCAN CHANNELS 2!1!1,
2. Enter the rest of the channels (Row 2 Column 2
through Row 2 Column 5)using the channel range
entry method:
A. Press “2”
Enters slot
B. Press “2”
Enters row of first channel
C. Press “2”
Enters column of first chan-
nel
D. Press “-”
Enters range designator and
slot
E. Press “2”
Enters row of last channel
F. Press “5”
Enters column of last channel
G. Press “
” or
Enters channel list terminator
“ENTER”
(comma)
SCAN CHANNELS 2!1!1, 2!2!2-2!2!5,
Note: The single channel entry method demonstrated
in step 1 could have been used to enter the channels for
row 2.
Step 4. Configure channel layer of scan.
NOTE
When RESET was performed (Step 1),
the channel layer was reset to MANU-
AL channel scanning. In MANUAL,
the first press of STEP will take the in-
strument out of the idle state and each
subsequent press will manually step
through the scan. Go ahead and try it.
When finished, press OPEN ALL to
place the Model 7001 in the idle state.
For this example, the channel scan rate will be con-
trolled by the timer. In general, every time the timer
times out, the next channel in the scan will be selected.
Since the Model 7001 was RESET, only CHANNEL
SPACING needs to be programmed.
Configure the channel layer as follows:
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Summary of Contents for 7001
Page 284: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 285: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 286: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 287: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 288: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 293: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 294: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 295: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 296: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 299: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 300: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 301: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 304: ......
Page 305: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 308: ......
Page 309: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 312: ......
Page 313: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 315: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...