4-25
4.6.2
GPIB
The GPIB menu is used for the following operations:
•
Check and/or change the address of the IEEE-488
bus.
•
Display the status byte of the IEEE-488 bus.
Perform the following steps to display the GPIB menu:
1. Display the MAIN MENU (see paragraph 4.6, Gen-
eral rule #1).
2. Use the cursor keys (
and
) to place the cursor
on GPIB, and press ENTER to display the follow-
ing menu:
GPIB SETUP MENU
ADDRESS
STATUS
Table 4-3
Default conditions
Item
Factory default
RESET
Channel status
All open
All open
Channel list
Cleared
No change
Scan list
No change
No change
GPIB address
No change
No change
Digital I/O output level
High
High
Language
No change
No change
Status messages
Off
Off
# of poles
No change
No change
Card pair
Off
Off
Delay
0 sec
0 sec
Channel spacing
Manual (STEP key)
Manual (STEP key)
Number of channels
Scan list length
Scan list length
Channel trigger control
Source
Source
Scan spacing
Immediate
Immediate
Number of scans
Infinite
Infinite
Scan trigger control
Acceptor
Acceptor
Arm spacing
Immediate
Immediate
Arm count
1
1
Arm trigger control
Acceptor
Acceptor
Single channel
Off
Off
Restricted channels
No change
No change
Break Before Make
On
On
Note: “No change” indicates that the currently set conditions are not affected.
ADDRESS
Use this menu item to check and/or
change the IEEE-488 address. At the factory the ad-
dress is set to 7, but can be changed to any value from
0 to 30. To select ADDRESS, place the cursor on AD-
DRESS and press ENTER. If the address is currently set
to 7, the following message will be displayed:
ADDRESS = 07 (0-30)
1. To retain the displayed address, press ENTER or
EXIT. The instrument will return to the GPIB SET-
UP MENU.
2. To change the address, simply key in a valid value
(0 to 30) and press ENTER. The instrument will re-
turn to the GPIB SETUP MENU.
STATUS
Use this menu item to display the IEEE-488
status byte. Refer to Section 4 (IEEE-488 Reference) for
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Summary of Contents for 7001
Page 284: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 285: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 286: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 287: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 288: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 293: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 294: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 295: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 296: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 299: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 300: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 301: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 304: ......
Page 305: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 308: ......
Page 309: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 312: ......
Page 313: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 315: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...