DC CURRENT UNCERTAINTY
= ± [ (ppm reading)
×
(measured value) + (ppm of range)
×
(range used) ] / 1,000,000.
% ACCURACY
= (ppm accuracy) / 10,000.
5PPM OF RANGE
= 10 counts at 6
½
digits.
DC IN-CIRCUIT CURRENT
The DC in-circuit current measurement function allows a user to measure the current through a wire or a circuit board trace without breaking the
circuit.
When the In-Circuit Current Measurement function is selected, the 2002 will first perform a 4-wire resistance measurement, then a voltage mea-
surement, and will display the calculated current.
TYPICAL RANGES:
Current:
100µA to 12A.
Trace Resistance:
1m
Ω
to 10
Ω
.
Voltage:
±200mV max. across trace.
Speed:
4 measurements/second at 1 power line cycle.
Accuracy:
±(5% +500µA). For 1 power line cycle, autozero on, 10-
reading digital filter, T
CAL
±5°C, 90 days, 1 year or 2 years.
AC AMPS
ACI ACCURACY
1,2
90 Days, 1 Year or 2 Years, T
CAL
±5°C, for 5% to 100% of range, ±(% of r % of range)
RANGE
20Hz–50Hz
50Hz–200Hz
200Hz–1kHz
1kHz–10kHz
10kHz–30kHz
3
30kHz–50kHz
3
50kHz–100kHz
3
200 µA
0.35 + 0.015
0.2
+ 0.015
0.4 + 0.015
0.5 + 0.015
2 mA
0.3 + 0.015
0.15 + 0.015
0.12 + 0.015
0.12 + 0.015
0.25 + 0.015
0.3 + 0.015
0.5 + 0.015
20 mA
0.3 + 0.015
0.15 + 0.015
0.12 + 0.015
0.12 + 0.015
0.25 + 0.015
0.3 + 0.015
0.5 + 0.015
200 mA
0.3 + 0.015
0.15 + 0.015
0.12 + 0.015
0.15 + 0.015
0.5 + 0.015
1
+ 0.015
3
+ 0.015
2
A
0.35 + 0.015
0.2
+ 0.015
0.3 + 0.015 0.45 + 0.015
1.5 + 0.015
4
+ 0.015
1. Specifications apply for sinewave input, AC+DC coupling, 1 power line cycle, autozero on, digital filter off, following 55-minute warm-up.
2. Add 0.005% of range uncertainty for current above 0.5A rms for self-heating.
3. Typical values.
AC CURRENT UNCERTAINTY
= ±[ (% of reading)
×
(measured value) + (% of range)
×
(range used) ] / 100.
PPM ACCURACY
= (% accuracy)
×
10,000.
0.015% OF RANGE
= 30 counts at 5
½
digits.
FREQUENCY COUNTER
FREQUENCY/PERIOD INPUT CHARACTERISTICS AND ACCURACY 90 Days, 1 Year or 2 Years
MINIMUM SIGNAL LEVEL
2
ACCURACY
FREQUENCY
PERIOD
RESO-
1HZ–
MAXIMUM
TRIGGER
±(% of
RANGE
1
RANGE
LUTION
1MHz 1–5MHz
5–15MHz
INPUT
LEVEL
reading)
AC Voltage Input
1Hz–15 MHz
67 ns – 1 s
5 digits
60 mV
60 mV
350 mV
1100 V pk
1
0–600V
0.03
AC Current Input
1Hz– 1 MHz
1 µs – 1 s
5 digits
150 µA
1 A pk
0–600mA
0.03
1. Subject to 2
×
10
7
V•Hz product (for inputs above 20V).
2. Valid for the lowest range. For each range increase, multiply these numbers by 10.
TIME BASE:
7.68MHz ± 0.01%, 0°C to 55°C.
READING TIME:
420ms maximum.
VOLTAGE INPUT IMPEDANCE:
1M
Ω
±2% with <140pF.
TRIGGER LEVEL ADJUSTMENT:
Trigger level is adjustable in 0.5% of range steps to ±60% of range in real-time using the up and down
range buttons.
FREQUENCY RANGING:
Autoranging from Hz to MHz.
FREQUENCY COUPLING:
AC + DC or AC only.
Specs and Accessories
4-5
MEASUREMENT RANGE CHART
10
Ω
1
Ω
100m
Ω
10m
Ω
1m
Ω
1m A 10mA100mA 1 A
10 A
10 0 A
100
µ
A
Measured Current
Specified
Measurement
Range