Voltage measurements
Assuming “BENCH” reset conditions, the basic procedure is as follows:
1.
Connect test leads to the INPUT HI and LO terminals. Either the front or rear inputs can
be used; place the INPUTS button in the appropriate position.
2.
Select the measurement function by pressing DCV or ACV.
3.
The AUTO annunciator indicates that autoranging is enabled. If you want manual rang-
ing, use the RANGE
▲
and
▼
keys to select a measurement range consistent with the
expected voltage.
4.
Connect test leads to the source as shown in Figure 4.
CAUTION
Do not apply more than 1100V peak or 2
×
10
7
V•Hz to the input, or instru-
ment damage may occur.
5.
Observe the display. If the “Over o w” message is displayed, select a higher range until
a normal reading is displayed. Use the lowest possible range for the best resolution.
6.
Take a reading from the display.
2001 MULTIMETER
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
EDIT
ERR
REM
TALK
LSTN SRQ
REAR
REL
FILT MATH
4W
AUTO
ARM
TRIG
SMPL
Caution: Maximum Input = 1100V peak, 2 x 10
7
V•Hz
AC Voltage
Source
Model 2002
Input Impedence = 1M
Ω
shunted by <140pF
Figure 4
DC and AC
voltage measurements
2001 MULTIMETER
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
EDIT
ERR
REM
TALK
LSTN SRQ
REAR
REL
FILT MATH
4W
AUTO
ARM
TRIG
SMPL
Model 2002
Caution : Maximum Input = 1100V peak
DC Voltage
Source
Input Resistance = 10M
Ω
on 1000V and 200V ranges.
> 100G
Ω
on 20V, 2V and 200mV ranges.
= 1M
Ω
on DCV peak spikes measurement.
1-10