Current measurements
Assuming “BENCH” reset conditions, the basic procedure is as follows:
1.
Connect test leads to the AMPS and INPUT LO terminals. Either the front or rear inputs
can be used; place the INPUTS button in the appropriate position.
2.
Select the measurement function by pressing DCI or ACI.
3.
The AUTO annunciator indicates that autoranging is enabled. If you want manual rang-
ing, use the RANGE
▲
and
▼
keys to select a measurement range consistent with the
expected current.
4.
Connect test leads to the source as shown in Figure 5.
CAUTION
Do not apply more than 2.1A, 250V to the input, or the amps fuse will blow.
5.
Observe the display. If the “Over o w” message is displayed, select a higher range until
a normal reading is displayed. Use the lowest possible range for the best resolution.
6.
Take the reading from the display.
2001 MULTIMETER
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
EDIT
ERR
REM
TALK
LSTN SRQ
REAR
REL
FILT MATH
4W
AUTO
ARM
TRIG
SMPL
Model 2002
Caution: Maximum Input = 2.1A
Current
Source
Figure 5