:FREQuency
Path to configure FREQ:
:COUPling <name>
Specify input coupling:
AC
or
DC
.
:REFerence <n>
Specify reference (rel): 0 to 15e6.
:STATe <b>
Enable or disable reference (rel).
:ACQuire
Use input signal as reference (rel).
:DIGits <n>
Specify measurement resolution: 4 or 5.
:SOURce <name>
Select source:
CURRent
or
VOLTage
.
:THReshold
Path to configure threshold:
:CURRent
Path to set current threshold:
:RANGe <n>
Specify range: 0 to 1.
:LEVel <n>
Specify level: ±60% of max signal level.
:VOLTage
Path to configure voltage threshold:
:RANGe <n>
Specify range: 0 to 1000.
:LEVel <n>
Specify level: ±60% of max signal level.
:TTL
Set threshold to TTL level.
:TEMPerature
Path to configure TEMP (RTD):
(DCV commands)
DCV commands (except
:RANGe
) apply.
:TRANsducer <name>
Select transducer:
RTD
,
FRTD
or
TCouple
.
:RTD
Path to configure RTD measurements:
:TYPE <name>
Select parameters:
PT385
,
PT3916
,
PT100
,
D100
F100
, or
USER
.
:ALPHa <NRf>
Specify alpha constant for USER: 0 to 0.01.
:BETA <NRf>
Specify beta constant for USER: 0 to 1.
:DELTa <NRf>
Specify delta constant for USER: 0 to 5.
:RZERo <NRf>
Specify rzero constant for USER: 0 to 1000.
Notes:
1.
Resolution (
:DIGits
) parameters:
All functions except FREQ and TEMP — 3.5 to 8.5.
FREQ— 4 to 5.
TEMP — 4 to 7.
2.
An advanced filter is available for DCV, ACV, DCI, ACI,
Ω
2, and
Ω
4. Refer to the User’s
Manual to use the advanced filter.
3-14
IEEE-488 Bus Operation