Kegel8 ULTRA User Manual Download Page 22

Kegel8

®

 Ultra Operation Manual

22

Specifi cations

STIM

1. 

Dual channel: individually isolated circuits.

2. 

Amplitude: 0-90 mA into 500 Ohm load; indication only. Actual mA  

 

will tend to be less than indicated due to electrode impedance: at  

 

1000 Ohms load (Electrodes in poor condition) the maximum will be  

 

limited to 86 mA, at 1500 Ohms load the maximum will be limited to  

 

65 mA.

3. 

Type: Constant Current, 

 

maximum output voltage 180 Volts +10 / -30 Volts

4. 

Waveform: Symmetrical, rectangular bi-phasic with zero DC current.

5. 

Selectable pulse width: 50 µS – 330 µS [10% accuracy].

6. 

Pulse Rate selection: in the continuous mode 2 – 100 Hz 

 

[5% accuracy].

7. 

Time duration of the treatment selectable: 1 minute to 90 minutes.

8. 

Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2 volts  

 

the battery symbol will fl ash on/off  once every second.

9. 

Open Electrode Detect: If an open circuit is detected at the 

 

output of channel A or B the output current will be reset at zero.

10. 

Ramp up time 0.1 - 9.9 seconds.

11. 

If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not  

 

turn on. 

 

Expected average battery set life [of standard 800 mAh, alkaline]:

 

19h.

Expected service life: 

5 years. Careful use and maintenance extends the life of the unit over the 
service life limit.

Environmental Conditions for use:

 

+5 to +40 degrees Centigrade.  15-93% Humidity.

Environmental conditions for storage & transport: 

-25 to +70 degrees Centigrade, 15-93% Humidity.

Weight:

 180g with battery.

Summary of Contents for ULTRA

Page 1: ...Kegel8 Ultra Operation Manual 1...

Page 2: ...r when you report a technical fault or claim a warranty return Manufacturer s serial number of the unit Present it together with LOT number when you report a technical fault or claim a warranty return...

Page 3: ...arger Never connect the Kegel8 Ultra directly to a battery charger or to any other mains powered equipment We advise not to use Ni Cad rechargeable batteries Patient Electrodes are for single patient...

Page 4: ...Instructions 9 Setting up your Customized Programme 11 Lock Button 12 Continence Treatment Programmes 13 Electrode Types Tips 19 Care Maintenance Accessories and Disposal 20 Indications for use 21 Sp...

Page 5: ...l8 Ultra is a dual channel device combining several treatment programmes into one unit Neuromuscular Stimulation is increasingly understood by Therapists and Doctors There is a better understanding of...

Page 6: ...lly dangerous equipment Do not place electrodes Over carotid sinus nerves Over larynx or trachea Inside mouth Over the area of the heart unless so advised by your Doctor On your facial area unless und...

Page 7: ...me SET button Reduces the intensity mA to zero and pauses the programme if a programme is running and moves the phase one step forward Displays the menu for programmes PC1 PC3 and allows the parameter...

Page 8: ...Time of current phase Number of phases Phase mode Work Rest Alt Cont Mod 3 Hz 250 uS PC1 5 WK Min 4SEC Programme Total programme time adds cumulatively as time added to phases Set Work Rest Ramp Delay...

Page 9: ...e to each of the pin connectors at the end of the lead wire The red and black polarity are good either way Some probes have a long wire built in with the direct socket to the unit 5 Make sure the Vagi...

Page 10: ...eturning to storage in the plastic bag Low Battery Indicator When the battery power is low the low battery indicator will appear on the screen shown in the diagram on page 7 When the battery indicator...

Page 11: ...est seconds between 2 and 99 seconds 9 Press CH A RP will appear and ash press CH B to set the ramp seconds between 0 1 and 9 9 seconds 10 Press CH A AL or SY will appear and ash press CH B to select...

Page 12: ...up correctly 2 Remove the battery cover and using a thin rod gently press on the lock button as shown in the diagram on page 6 until you hear a double beep The unit is now locked and cannot be altered...

Page 13: ...on S 150 Ramp up time secs 1 0 Ramp down time secs 0 Work time secs Cont Rest time secs 0 Alternating Synchronous Overall time 20 min Programme P02 Urge Incontinence Phase 1 Phase 2 Phase 3 Phase 4 Ph...

Page 14: ...p time secs 1 0 Ramp down time secs 0 Work time secs 6 Rest time secs 15 Alternating Synchronous Overall time 20 min Programme P04 Stress Incontinence 2 Phase 1 Phase 2 Phase 3 Phase 4 Phase 5 Phase t...

Page 15: ...time secs 1 0 Ramp down time secs 0 Work time secs 5 Rest time secs 5 Alternating Synchronous Overall time 20 min Programme P06 Frequency Urge 2 Phase 1 Phase 2 Phase 3 Phase 4 Phase 5 Phase time min...

Page 16: ...ously with no rest period It is used in some countries where they have found continuous stimulation can work e ectively Programme P08 Lack of Sensitivity Phase 1 Phase 2 Phase 3 Phase 4 Phase 5 Phase...

Page 17: ...Ramp up time secs 0 8 0 8 0 8 0 6 0 8 Ramp down time secs 0 0 0 0 0 Work time secs 5 5 5 5 5 Rest time secs 5 5 5 6 7 Alternating Synchronous Overall time 45 min Programme P10 Building Up Endurance P...

Page 18: ...uration S 220 Ramp up time secs 1 2 Ramp down time secs 1 2 Work time secs 6 Rest time secs 10 Alternating Synchronous Overall time 20 min This programme helps to relax the pelvic muscle It may be use...

Page 19: ...y cleansing the skin with a swab impregnated with alcohol Clip away hairy skin using scissors don t use a razor to remove the hair The electrodes conductive material is water based If it becomes satur...

Page 20: ...ically for any discharge from the battery Remove battery completely from unit if not in use for any extended period of time typically one week Low battery indicator of 6 9 volts shown on LCD display W...

Page 21: ...they drop onto the oor debris will adhere to conductive gel rendering the electrodes ine ective Electrode life can be considerably reduced by The type and condition of the skin Deep seated moisturiser...

Page 22: ...accuracy 7 Time duration of the treatment selectable 1 minute to 90 minutes 8 Low Battery Indicator If the battery goes below 6 9 volts 0 2 volts the battery symbol will ash on o once every second 9 O...

Page 23: ...RF emissions interference to be immune from e ects of interference produced by other equipment operating in their vicinity and damage due to electrostatic discharge all when operating in a typical do...

Page 24: ...Kegel8 Ultra Operation Manual 24...

Page 25: ...Kegel8 Ultra Operation Manual 25...

Page 26: ...commend that the next best position is to sit down and lean back slightly 5 If you think the probe itself is not working wash it and hold it using your rst nger and thumb or elbow crook to make a conn...

Page 27: ...for any customer service enquiries including the warranty returns Your invoice of purchase and or the rear cover of this manual should state the name and the contact details of your distributor For a...

Page 28: ...Kegel8 Ultra Operation Manual 28 Clinical References Please contact us for any clinical referencies of Kegel8 enquiries kegel8 com...

Page 29: ...Kegel8 Ultra Operation Manual 29 Notes...

Page 30: ...Kegel8 Ultra Operation Manual 30...

Page 31: ...Kegel8 Ultra Operation Manual 31...

Page 32: ...A Document revision info Kegel8 Ultra manual English OPH200 OM EN13 23 01 2019 Savantini Ltd Savantini House Foster Street Stoneferry Road Hull HU8 8BT England Any queries should be addressed to Helpl...

Reviews: