5. Dry the disinfected Tonometer prism
with a clean soft non-shedding cloth.
6. Store the Tonometer prism in a
suitable container ready for use.
Safely dispose of the disinfectant fluids used.
Do not disinfect using the following: Alcohol, Acetone, UV radiation,
Sterilisation, Immersion in fluid for more than one hour, Temperatures
greater than 60°C.
4. TONOMETER PRISM FIELD CHECK
Check the Tonometer prism under the Slit lamp and ensure there are no
cracks / chips. the chemical used in the diagnosis process (for example:
Fluorescein) will get into the cracks and will show up if observed under
the slit lamp. Do not use it if there is any sign of cracks / chips.