D-KAT Z-Type Tonometer by KEELER
3.2 DISINFECTING THE TONOMETER PRISMS
Always disinfect the Tonometer prisms before use. Hand hygiene must be
considered to prevent any contamination.
1. Carefully remove the Tonometer prism
from the prism holder.
2. Wash the Tonometer prism under cold
running water for approximately 1
minute, to ensure the Tonometer prism
is physically clean before exposed to
the disinfection process.
1 min
3. Immerse the Tonometer prism in the
disinfectant fluid. Types of disinfectant
fluid vary.
4. Rinse the disinfectant from the prism
in running water for between 10 and
30 minutes.
10 mins
10-30 mins
Please follow disinfectant solution guidelines
for instructions, concentration and time of
immersion. (For example: Pantasept – 3%
aqueous solution for 10 minutes, Hydrogen
Peroxide 3% aqueous solution for 10 minutes,
Sodium Hypochlorite, 10% aqueous solution for
10 minutes etc.).