D-KAT Z-Type Tonometer by KEELER
6. Once the spirit level indicates the ZKAT is level, lock the ZKAT in position by tightening
the large silver screw provided.
7. Check that the ZKAT is within 1-degree in the roll orientation. If the ZKAT is in more
than 1-degree roll adjust the mounting of the slit lamp.
8. Re-check ZKAT prism is still concentric to the viewing optics, re-adjust if required.
7.2 PREPARING THE PATIENT
• Use an appropriate topical anaesthetic to numb the cornea.
• Place a strip of fluorescein-soaked paper near the outer canthus in the lower conjunctival
Screw to elevate
the ZKAT
Screw to decline
the ZKAT