7
EN
D-KAT Z-Type Tonometer by KEELER
Batteries
• Do not use a battery that is deformed, leaking, corroded or visually damaged. Handle
a damaged or leaking battery with care. If you come into contact with electrolyte,
wash exposed area with soap and water. If it contacts the eye, seek medical attention
immediately.
• Ensure battery orientation is correct, or personal injury / damage to equipment may occur.
• Do not mix battery types.
• Do not attempt to charge non-rechargeable batteries.
• Dry cell batteries should be removed if your instrument is not to be used for long periods.
• Do not dispose of battery in fire, puncture or short circuit.
• Dispose of batteries in line with local environmental regulations.
• Tape over battery contacts to avoid short circuiting during disposal.
• After removal of the battery do not touch the battery contacts and the patient
simultaneously.
2.2 CONTRAINDICATION
There is no restriction to patient population this device can be used with and there are no
known contraindications associated with the device.
• Do not use contact tonometry on those with ruptured globe, corneal lesions, ulcer or
epithelial defects and patients that are not suitable for anaesthesia administration.
• Take extra care with patients with ocular infection to prevent transmission.
3. CLEANING AND DISINFECTION INSTRUCTIONS
3.1 CLEANING TONOMETER BODY
Only manual non-immersion cleaning as described should be used for this instrument. Do not
autoclave or immerse in cleaning fluids. Always disconnect power supply from source before
cleaning.
1. Wipe the external surface with a clean absorbent, non-shedding cloth dampened with
de-ionised water / detergent solution (2% detergent by volume) or water / isopropyl
alcohol solution (70% IPA by volume). Avoid optical surfaces.
2. Ensure that excess solution does not enter the instrument. Use caution to ensure cloth is
not saturated with solution.
3. Surfaces must be carefully hand-dried using a clean non-shedding cloth.
4. Safely dispose of used cleaning materials.