KCI ActiV.A.C. Owner'S Service Manual Download Page 8

6

Therapy Unit

1.  Remove the unit from the plastic and/or fabric carrying case, as 

applicable.

2.  Ensure therapy unit is unplugged from power supply.
3.  Remove canister seals from unit.  Wipe/clean seals with antiseptic.  

Ensure the seal cavities and adjacent surfaces are clean of any foreign 
material.  Replace seals when finished.

Canister Seals

4.  Remove the Exhaust Filter.  Discard the filter using local protocols.  

Moisten a cotton tipped applicator with antiseptic; thoroughly wipe the 
interior of the filter compartment.

The Exhaust Filter is single use only.

5.  Wipe down the back of the unit with antiseptic.

Do not allow excess cleaning fluid into the areas surrounding 
the Touch Screen Gasket.

6.  When wiping the front of the unit, do not allow cleaning fluid to seep 

into any gaps in the Touch Screen Gasket.  Three such typical areas are 
pointed out in the illustrations below.

Cleaning fluids allowed to seep past the gasket may damage 
the Touch Screen to the point of requiring replacement.

Exhaust Filter

Summary of Contents for ActiV.A.C.

Page 1: ...ActiV A C Therapy Unit Owner Service Manual 418474 Rev A February 2018 Important Document File in your maintenance records...

Page 2: ......

Page 3: ...sure Checks 13 Set up 13 25 mmHg Test 14 125 mmHg Test 14 200 mmHg Test 15 125 mmHg Test 16 Charge Battery and 6 Hour Unit Verification Test 17 Verify Time and Date 20 Alarm Tests 20 Leak Alarm 20 Blo...

Page 4: ...any service procedures on this equipment In order for KCI products to perform properly KCI recommends the following Failure to comply with these conditions will void any applicable warranties Assembly...

Page 5: ...Unit for use includes unpacking inspection for any damaged or missing parts and an initial round of service procedures to ensure the battery is fully charged and the unit is ready for patient use The...

Page 6: ...tton Battery Charging Indicator LED USB Data Port Manufacturer Service Port Tripping Hazard No bathing or showering Tripping Hazard No bathing or showering Mains Power Cord Power Supply Power Cord War...

Page 7: ...l protective equipment PPE and hand hygiene protocols in accordance with the following standards 29 CFR 1910 1030 OSHA Bloodborne Pathogens Standard MMWR October 2002 51 No RR 16 Guidelines for Hand H...

Page 8: ...otocols Moisten a cotton tipped applicator with antiseptic thoroughly wipe the interior of the filter compartment The Exhaust Filter is single use only 5 Wipe down the back of the unit with antiseptic...

Page 9: ...unplugged from the therapy unit and or any power source 2 Wipe the power supply and cord with antiseptic Allow to air dry 3 Inspect the power supply brick cords and connectors for damage and cracked...

Page 10: ...C Canister KCI P N M8275058 two 2 each These are normal ActiV A C canisters that can be reused for up to 6 months as a test canister KCI recommends marking the canister with Test Canister Not For Huma...

Page 11: ...t filter This procedure can be skipped if the unit is new and has not been previously placed with a patient 1 Insert a new exhaust filter 2 Ensure the filter is fully seated 3 With the Exhaust Filter...

Page 12: ...n external computer The ActiV A C unit exports a file folder that uses the unit serial number as its name The folder contains a Comma Separated Values CSV file named by date YYYYMMDD csv 14 For the in...

Page 13: ...battery 1 Verify that the Battery Change Due date on the Battery History label is at or exceeds the current date 2 Remove the Battery History Label 3 Using a 1 Phillips screwdriver remove the two 4 40...

Page 14: ...using ensuring the wiring harness is on the front side of the battery and NOT behind and is otherwise not tangled or obstructing fitting the battery fully into the battery compartment 14 Press the Pow...

Page 15: ...e Valve Hygroscopic Filter connected to Pump Port Power On Off Switch Hygroscopic Filter connected to Pump Port 4 Ensure the tubing with the hygroscopic filter is connected to the pump port and the pl...

Page 16: ...Pressure Exit _ 25 mmHG 7 Press OK 8 Start therapy by pressing On Off the Therapy Start screen will appear Continuous Therapy Off 0 mmHg Confirm OK Back Pressure 25 mmHg Mode Continuous Intensity Low...

Page 17: ...l manometer and the displayed pressure on the ActiV A C Touch Screen 10 The digital manometer should stabilize and indicate the pressure displayed touch screen 10 mmHg 11 Press Exit one 1 time 12 Pres...

Page 18: ...Hg Pressure Exit _ 125 mmHG 3 On the Confirm screen ensure that settings are Pressure 125 mmHg Mode Continuous Intensity Low 4 Press OK 5 Observe the digital manometer and the displayed pressure on th...

Page 19: ...ll socket 5 Verify the Battery Charging Indicator LED is lit Battery charging in process Battery charging complete An amber light means the battery is charging A green light means the battery is fully...

Page 20: ...ge 28 Continuous Therapy Off 0 mmHg Pressure Exit _ 125 mmHG Continuous Therapy Off 0 mmHg Confirm OK Back Pressure 125 mmHg Mode Continuous Intensity Low Clinician Mode 6 Aug 2009 15 06 12 Attach a r...

Page 21: ...ime period is achieved 26 Press On Off to stop therapy 27 At the end of the six hour verification period carefully check the unit for Control buttons not responding Excessively hot unit No alarms are...

Page 22: ...4 Set the correct date and time 5 Press Exit two times to return to the Clinician Home Screen Exit Set Time Date Year Minute Hour 24 Day Month _ _ _ _ _ Alarm Tests The purpose of these tests is to e...

Page 23: ...n press On Off 4 From the Therapy Start screen press Exit to return to the Clinician Home screen 5 The Blockage Alert should activate in approximately two minutes or less 6 Press Exit to return to the...

Page 24: ...er Not Engaged alarm should sound in approximately thirty 30 seconds or less 9 Ensure that the alarm sounds with three 3 beeps 10 Press Reset to return to the Clinician Home Screen 11 Continue on to C...

Page 25: ...e ActiV A C Therapy Unit still powered off set the unit in a secure place where it can be left for a minimum of six hours 2 Plug the power supply cord into the ActiV A C unit 3 Plug the mains power co...

Page 26: ...try appropriate mains power cord Preparation For Use 1 Carefully wrap the power cords into a bundle and place the power supply into a clear plastic bag 2 Position the ActiV A C Therapy Unit in the fab...

Page 27: ...the fabric carrying case flap to provide additional accidental impact protection to the Touch Screen when the flap is properly closed 2 Insert the ActiV A C Therapy Unit Quick Reference Guide and V A...

Page 28: ...y Voltage Input 100 240 VAC Frequency 50 60Hz Amperage 0 72A 115VAC External Power Supply Output 12V 3 3 A Patient Enclosure Leakage Current 100 Microamps Battery Information Type Lithium ion Li ion P...

Page 29: ...ning Label 2 required 340168 ActiV A C Battery History Label 340171 ActiV A C Logo Label 340225 ActiV A C North American Mains Power Cord 340226 ActiV A C Power Supply 340367 ActiV A C Compliance Labe...

Page 30: ...cal and Electronic Equipment Directive 2002 96 EC At the end of useful life dispose of all waste according to local requirements or contact your local KCI subsidiary or agent for advice Not protected...

Page 31: ...ention and Control process pg 5 Inspect Unit For Damage pg 8 Blockage Alert pg 21 Canister Full Therapy Interrupted pg 22 Recharge Battery pg 23 Contact KCI see page 28 Fail Fail Fail Fail Fail Fail F...

Page 32: ...Manufactured for KCI USA Inc 12930 IH10 West San Antonio Texas 78249 USA www acelity com...

Reviews: