13
2. Connect the PVC test fixture to the ActiV.A.C.
®
unit as illustrated below.
3. Ensure the calibrated manometer is connected as illustrated
and powered on.
Connected to
Power Supply
ActiV.A.C.
®
Unit
PVC Test Fixture
Customer Supplied
Calibrated Manometer
Needle Valve
Hygroscopic Filter
connected to Pump Port
Power On/Off
Switch
Hygroscopic Filter
connected to Pump Port
4. Ensure the tubing with the hygroscopic filter is connected to the pump port and the plain tubing is connected to the
wound port.
5. Ensure the ActiV.A.C.
®
Power Supply is plugged into mains power and is also plugged into the ActiV.A.C.
®
unit.
Power the unit on by pressing the power on/off switch for two seconds.
Plain tubing connected
to Wound Port