The purpose of this procedure is to verify the unit displays the correct time
and date and how to set it if it is incorrect.
1. With the unit powered on and the Clinician Home Screen visible,
observe the time and date in the upper right-hand corner of the screen.
If the time and date are correct, proceed to ‘Alarm Tests’, otherwise go
to Step 2.
2. Press ‘Utilities’.
3. Press ‘Time/Date’.
4. Set the correct date and time.
5. Press ‘Exit’ two times to return to the Clinician Home Screen.
Exit
Set Time/Date
Year
Minute
Hour(24)
Day
Month
+
+
+
+
+
_
_
_
_
_
Alarm Tests
The purpose of these tests is to ensure the ActiV.A.C.
®
Therapy Unit
correctly detects, displays and sounds alert and alarm conditions.
Ensure that the ActiV.A.C.
®
Therapy Unit is NOT plugged
in to the ActiV.A.C.
®
power supply during the alarm test
procedures. Ensure the tests are run without interruption as
an additional battery capacity check.
Leak Alarm
1. Ensure the ‘test canister’ is still securely attached to the ActiV.A.C.
®
unit. Remove the tubing cap from the canister tubing and ensure the
tubing clamp is completely open.