22. Start therapy by pressing ‘On/Off’, the ‘Therapy Start’ screen will appear.
23. Ensure the Leak Rate indicator is green.
24. Press ‘Exit’ to return to the Clinician Mode Home screen.
Continuous
Therapy On
125mmHg
On/Off
Clinician Mode
6 Aug 2009
15:06
Utilities
Therapy
25. KCI recommends starting a timer to ensure the six hour time period is
achieved.
26. Press ‘On/Off’ to stop therapy.
27. At the end of the six hour verification period, carefully check the unit
for:
• Control buttons not responding
• Excessively hot unit
• No alarms are displayed
28. Verify that the Battery Charging Indicator LED is lit and showing green.
Continuous
Therapy Off
0mmHg
On/Off
Clinician Mode
6 Aug 2009
15:06
Utilities
Therapy
29. Unplug the power supply cord from the ActiV.A.C.
®
unit.
30. Verify the ‘battery full’ icon appears.
31. Should the unit fail any of the above criteria, contact KCI (see page 28).
32. With the ‘Charge Battery and Six Hour Verification Test’ complete,
continue on to ‘ Verify Time and Date‘.