22
Canister Not Engaged and Check Battery Level
1. From the Clinician Home Screen press ‘On/Off’ to stop therapy.
2. Release the canister from the unit by pressing the Canister Latch
Release.
3. From the Clinician Home screen, press ‘On/Off’.
4. From the ‘Therapy Start’ screen, press ‘Exit’ to return to the Clinician
Home screen.
5. While this test is running, check the status of the battery icon on the
Clinician Home Screen.
6. If the battery icon is yellow as illustrated, battery capacity is not
sufficient for normal operation. Cancel this test and go to the ‘Change
Battery’ procedure and change the battery. After changing the battery,
repeat the procedures and tests from that point.
Continuous
Therapy Off
0mmHg
Audio Pause
Reset
WARNING: Ensure canister is in place.
Press ‘?’ for more information.
Canister Not Engaged
7. If the battery icon remains green, continue with this test.
Continuous
Therapy Off
0mmHg
Audio Pause
Reset
WARNING: If canister is full, replace and press ‘Reset’.
If not full, press ‘Cancel’. Press ‘?’ for more information.
Canister Full
Therapy Interrupted
Cancel
8. The Canister Not Engaged alarm should sound in approximately thirty
(30) seconds or less.
9. Ensure that the alarm sounds with three (3) beeps.
10. Press ‘Reset’ to return to the Clinician Home Screen.
11. Continue on to ‘Canister Full Therapy Interrupted’ alarm if the unit has
correctly displayed this ‘Canister Not Engaged’ alarm, otherwise contact
KCI (see page 28).
Continuous
Therapy On
0mmHg
On/Off
Clinician Mode
6 Aug 2009
15:06
Utilities
Therapy
Continuous
Therapy On
0mmHg
On/Off
Clinician Mode
6 Aug 2009
15:06
Utilities
Therapy
Canister Full Therapy Interrupted
1. Remove the canister from the unit and replace it with a canister that
has the canister sensing port covered with a piece of impermeable tape
cut and placed as shown.
2. Ensure the canister latches into place.
3. From the Clinician Home screen, press ‘On/Off’.
4. From the ‘Therapy Start’ screen, press ‘Exit’ to return to the Clinician
Home screen.
5. The ‘Canister Full Therapy Interrupted’ alarm should sound in
approximately ninety (90) seconds or less.
6. Ensure that the alarm sounds with three (3) beeps.
7. Press ‘Reset’ to return to the Clinician Home Screen.
8. Remove this test canister from the unit.
9. Continue on to ‘Final Settings’ if the unit has correctly displayed this
‘Canister Full Therapy Interrupted’ alarm, otherwise contact KCI (see
page 28).