Chapter 5. Preparing for Calibration
5.1 Selecting LTR-150 Function
The LTR-150 is supplied standard with a dry block insert. To utilize the optional micro bath or surface
temperature functions, the inserts must first be changed and the appropriate Function defined via the
touchscreen.
Insert the appropriate insert based on the calibration function required.
Each measuring insert has its own characteristic values, which are defined at the factory. These characteristic
values are influenced by different factors, such as: diameter, type and material of the measuring insert,
number and diameter of the bore holes or the physical properties of the calibration liquid.
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in serious burn
injuries.
Never touch the metal block, the tank, the adapter sleeve or the test sensors at temperatures above
35 °C or below 10 °C.
Allow the calibrator to cool to ambient before removal or insertion of any insert.
Always power off unit and clean the tank and the desired insert before insertion.