23
M5101-EN Rev. B |
IMPORTANT NOTICE!
Use only recommended silicone oil.
Always read the safety data sheet supplied with the silicone oil before using it.
Always ensure adequate ventilation when working with silicone oil, since hazardous
substances can be released.
Spilled or leaked silicone oil results in an extreme danger of slipping. Clean the affected
ranges by suitable means.
Since silicone oil is hygroscopic, always use the associated transport cover to close the
tank or the tub insert after use.
Installation:
Insert the tub insert in the tank using the insert exchange tool.
Place the magnetic stirrer in the tank / tub insert.
Insert the sensor cage.
Insert the test sensors in the sensor cage.
This accounts for the volume of the sensors to be tested.
Fill the calibration liquid in the tub insert to the fill level.
Leave sufficient reserve space for additional level rise.
If necessary, remove the test sensors again.
Screw the work cover onto the tank and if necessary insert the sensors through the work cover
into the tub insert.
Removal:
Let the calibrator cool before you remove the tub insert.
If necessary, remove the working cover or transport cover.
Pull the tub insert out of the metal block with the help of the
insert exchange tool.
Empty fluid.
Remove and clean the sensor cage, magnetic stirrer, and tub
insert.
Clean the metal block.
Chapter 5. Preparing for Calibration