Operating Instruction STERIclave B 2200 / 2200 P
6 Operation
|
6.6 Tracking the program sequence
6.6 Tracking the program sequence
Phase
Display/LED/
Acoustic signal
Time display
Pressure display
Temperature display
Program display
Approx. 10 seconds
Duration of selected
program
Pressure of selected
program
Temperature of se‐
lected program
Vacuum phase
5-10 minutes
Vacuum lights up
Remaining time
in phase
Current pressure
Current temperature
Sterilisation phase
Sterilize lights up
Remaining time
in phase
Current pressure
Current temperature
Decompression
Sterilize lights up
Remaining time
in phase
Current pressure
(falls back to zero)
Current temperature
(drops)
End of sterilisation
Sterilize flashes
Remaining time
in phase
Current pressure
Current temperature
Aspiration of residual
vapour
Drying lights up
Process duration
Current pressure
Current temperature
Drying
Drying lights up
Remaining time
in phase
Current pressure
Current temperature
Airing
Drying lights up
Remaining time
in phase
Current pressure
Current temperature
End of drying phase
(class B sterilisati‐
ons)
Ready and Sterilize
light up + acoustic
signal sounds
Total duration of pro‐
cess
Current pressure
Current temperature
End of drying phase
(Fast cycle and Spe‐
cial S4)
Ready lights up +
acoustic signal
sounds
Total duration of pro‐
cess
Current pressure
Current temperature
If the door is opened within half an hour of the drying phase finishing, the heating
will remain in the pre-heating phase.
50/78