Malfunction
Cause
Remedy
Sieves in suction connector
blocked.
▶ Exchange the sieve.
Foot switch for vacu-stop has been
pressed.
▶ Release the foot switch.
Suction machine not running.
▶ Turn on the suction machine.
▶ Check the suction machine fuse.
The amalgam separator does not
work correctly.
▶ See the operating instructions
for the amalgam separators.
The signal sounds continuously,
and the "Service" LED (yellow)
flashes.
Warning about the amalgam sepa‐
rator.
▶ See the operating instructions
for the amalgam separators.
The buzzer sounds every 10 sec‐
onds.
The Oxygenal container is empty. ▶ Fill the Oxygenal container.
(See care instructions.)
A beep sounds ten times.
The Oxygenal container is too full. ▶ Stop filling the Oxygenal con‐
tainer.
The Satelec Mini LED does not
work.
See also: Instructions for use for the
Satelec Mini LED
See also: Instructions for use for the
Satelec Mini LED
The patient chair does not move.
The safety shutoff is activated.
(The LED on the control panel flash‐
es.)
▶ Check the safety shutoff and
eliminate the reason for the
shutoff.
The patient chair does not move, or
only moves upward slightly.
The spittoon is swung toward the
patient chair (the safety shutoff is
activated).
▶ Swing the spittoon bowl into
resting position.
The "Device on" LED (green) flash‐
es.
Internal error.
▶ Call a technician.
The "Service" LED (yellow) flashes. Malfunction in the amalgam sepa‐
rator.
▶ Call a technician.
Emergency shut off of the bowl
valve (only when external suction is
installed)
▶ Call a technician.
The buzzer sounds every second,
and the "Device on" LED (green)
and "Service" LED (yellow) alter‐
nately flash.
Leakage water in the base of the
device (only when the DVGW water
block is installed).
▶ Shut off the water supply.
▶ Call a technician.
The LEDs of the keys "AP 0", " AP
1", " AP 2" and "SP" flash for 10 sec‐
onds after the patient chair has
been activated.
The position pickup is defective or is
incorrectly addressed.
▶ Call a technician.