3.8
Technical Data
Drilling template and setup plan
Drilling template (paper)
(Mat. no. 1.000.3001)
Page 001
Drilling template (sheet met‐
al)Mat. no. ). 10.002492
Layout plan (Mat. no. 1.000.3001)
Sheet 002
Electrical system
Electrical lead
3 x 1.5 mm
2
Free end above the floor
1 000 mm
Input voltage
100/110/120/130/220/230/240 V AC
Frequency
50 Hz
Input voltage set by the manufacturer
See rating plate
Max. power consumption (including the
KAVOLUX 1410) at 100 to 130 V
30 to 600 VA
Max. power consumption (including the
KAVOLUX 1410) at 220 to 240 V
30 to 1,000 VA
Customer fuse protection
Automat C16 or screw-plug fuse 10 A
Protective conductor above floor
see DIN VDE 0100-710, 1,000 mm
Heat emission at 100 to 130 V
97 to 2,106 kJ/h
Heat emission at 220 to 240 V
97 to 3,240 kJ/h
Mark of approval
CE / DVGW / VDE
Footswitch
IPX1 (moisture protection)
Water supply
Note
If the water is very hard (above 12 dH), a water softening device must be fitted in
the ion-exchange process.
Insufficient water hardness (below 8.4 dH) can promote the formation of algae.
National water regulations apply to the compact water block kit. The following also
applies: DIN EN 1717.
According to DIN EN 1717, each unit that is not listed by the DVGW in last the
provided with an upstream type AA, AB or AD a safety device. (The DVGW water
block kit and DVGW water bottle are certified; see the following list.)
When creating a water connection, prevent brackish water pools with standing water
(also in the house plumbing).
You can find additional information at www.dvgw.de
Instructions for use Status 1080 TM
3 Product description | 3.8 Technical Data
21/74
Summary of Contents for Status 1080 TM
Page 1: ...Instructions for use Status 1080 TM Always be on the safe side...
Page 77: ......
Page 78: ...1 000 6141 Fk 20100309 04 en...