Note
Take into account the additional measuring points of additional equipment such as
connection to equipment by a different manufacturer, treatment lamps, multimedia
system, etc.
▶ Disconnect L + N at the device side from the mains, or connect measuring line
Mat. no. 0.411.8811 to X 2.
▶ Check the EUL at the measuring points or with an EPA measuring line
Mat. no. 1.001.9904.
8.3
Measuring dummy patient leakage current
Threshold: < 5 mA
The equivalent unit leakage current (EUL) can be measured using an EPA meas‐
uring line (Mat. no. 1.001.9904) or at the falling measuring points on the patient
chair:
▪ Fuses F4, F7 and F10 and X
Instructions for use Status 1080 TM
8 Carrying out safety checks | 8.3 Measuring dummy patient leakage current
69/74