Removing the cannulas
▶ Remove the cannula with a slight amount of rotation while holding the handpiece
on the grip sleeve.
Setting cold light intensity
The cold light intensity is adjusted in the same manner as with the turbine.
Changing the high-pressure lamp
CAUTION
Injury hazard from hot valve body.
Burns from contact.
▶ Switch main device switch off.
▶ Let the lamp cool down after long use.
▶ Pull off the grip sleeve ④ together with the cannula from the valve body ①.
▶ Push the holder ③ forward, and pull the defective high-pressure bulb ② out of
the socket.
▶ Insert the new high-pressure lamp (Mat. no. 1.002.2928).
Check the position of the contacts.
▶ Pull back the holder ③ slightly.
▶ Push the grip sleeve and cannula on together until they audibly lock into place.
Instructions for use Status 1080 TM
4 Operation | 4.8 Using instruments
59/74