4.8.7
Setting the spray mist suction device and saliva ejector
CAUTION
Unintentionally activating the saliva ejector or spray fog suction.
Injury in the area of the mouth.
▶ When using the vacuum stop to interrupt suction, do not leave the saliva ejector
or spray fog suction in the patient's mouth.
Adjusting the suction strength with the slide
▶ Remove the saliva ejector or spray mist suction device from the holder.
The suction automatically turns on.
▶ To set the suction or block it, move the slides that are integrated in the conical
pieces of the saliva ejector and spray fog suction.
Start stop the suction with the vacuum stop (extra)
▶ Remove the saliva ejector or spray mist suction device from the holder.
The suction automatically turns on.
▶ To interrupt the suction, press and hold the "vacuum stop" button.
Assign the "Vacuum stop" key
▶ In service mode, (group 9), assign the "Vacuum stop" key.
The following assignments are possible:
▪ Simultaneously stop the spray mist suction device and saliva ejector
▪ Only stop the spray mist suction
Instructions for use Status 1080 TM
4 Operation | 4.8 Using instruments
60/74