Ambient conditions during operation
Temperature
°C
+10 to +35
Relative humidity
%
20 - 80
Air pressure
hPa
750 - 1060
Height above sea level
m
< 2000
Ambient conditions during storage and transport
Temperature
°C
-18 to +60
Humidity
%
10 - 95, non-condensing
Air pressure
hPa
500 - 1060
Classification
Medical Devices Directive (93/42/EEC): Class I
Laser class (unit) in accordance with EN 60825-1:2014: 1
Laser source
Laser class
3B acc. to IEC 60825-1:2014
Wavelength
nm
635
Power
mW
<10
Technical data of the RFID Module
Frequency
MHz
13.56
Modulation
ASK
Max. power
mW
400
Electromagnetic compatibility (EMC) interference emission measurements
High-frequency emissions in accordance with
CISPR 11 Group 1
Group 1, Class B
Interference voltage at the power supply
connection CISPR 11:2009+A1:2010
Compliant
Electromagnetic interference radiation CISPR
11:2009+A1:2010
Compliant
Electromagnetic compatibility (EMC) interference immunity measurements
cover
Immunity to interference, discharge of static
electricity
IEC 61000-4-2:2008
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Compliant