Steps
1. Place the imaging plate in the light protection cover into the patient's
mouth.
When doing this, make sure that the active side of the imaging plate
points towards the X-ray tube.
2. Set the exposure time and setting values on the X-ray unit (see section
"X-ray unit settings" on page 31).
3. Record an X-ray image.
The image data must be scanned within 30 minutes.
4.2.2.4 Preparing for scanning
About this task
CAUTION
Light erases the image data on the imaging plate
Never handle exposed imaging plates without the light protection cover.
Wear protective gloves.
Steps
1. Remove the imaging plate with the light protection cover from the
patient's mouth.
WARNING
Contamination of the unit
Clean and disinfect the light protection cover before removing the
imaging plate.
2. In the event of heavy soiling, e.g. from blood, dry clean the light
protection cover and protective gloves, e.g. wipe with a clean cellulose
cloth.
4 Usage
42 KaVo ProXam iP
Instructions for use