3.3 Commissioning
CAUTION
Short circuit can occur due to build up of condensation.
Do not switch on the unit until it has warmed up to room temperature and it is
dry.
The unit supports the following imaging programs:
•
Romexis
3.3.1 Network configuration
Various options are available for network configuration:
•
Automatic configuration via DHCP (default)
•
Automatic configuration via Auto-IP for direct connection of unit and
computer (default, in case DHCP is not available) with firmware version
1.3 or later: if no DHCP response, uses Auto-IP: 169.254/16
•
Manual configuration (recommended)
Configure the network settings of the unit using the KaVo Device Tool for
ProXam iP.
Check the firewall and release the ports, if applicable.
3.3.2 Network protocols and ports
Port
Purpose
Service
45123 UDP
45124 UDP
Unit recognition and
configuration
2006 TCP
Unit data
514
1)
UDP
Event protocol data
Syslog
1)
The port can vary depending on the configuration.
NOTE
When the unit is first connected to a computer, it applies the language and
time settings of the computer.
3.3.3 Configuring unit in KaVo Device Tool for ProXam iP
About this task
Configuration is performed using KaVo Device Tool for ProXam iP.
NOTE
The configuration with the KaVo Device Tool for ProXam iP should be
performed by qualified service technician only.
For more detailed instructions, see
KaVo ProXam iP Technician's
instructions.