▶ Clean the assistant element with a soft cloth and a mild, non-abrasive clean‐
ing solution.
2.5.2
Sterilisation assistant element
CAUTION
Damage caused by sterilisation.
▶ Do not sterilise the assistant element.
2.5.3
Preparing the connector for the suction hose
▶ Remove connectors ① and ② from the suction hoses.
Cleaning and disinfection of the suction hose connector
▶ Clean the disassembled parts under running water.
▶ Thermodisinfect the disassembled parts for at least 10 minutes at 93 °C (total
cycle: approximately 20 minutes)
Sterilising the suction hose connector
CAUTION
Damage caused by sterilisation.
▶ Do not sterilise the suction hose connector.
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant unit