Cleaning the kit for physiological saline solution
When using saline solution, all lines conducting the solution must be rinsed direct‐
ly after treating the patient.
▶ Draw the pump hose with the needle from the NaCl bag.
▶ Immerse the pump hose needle in a container with distilled water (min. 150
ml).
▶ Activate the pump, and pump distilled water completely through it.
Sterilisation kit for physiological saline solution
Note
The following parts can be subjected to sterilisation:
– Silicone hoses for saline solution and their connecting nipples
CAUTION
Damage caused by sterilisation.
▶ Do not sterilise instrument hoses and hose clips for saline hose.
Requirement
The hoses for the physiological saline solution are rinsed.
▶ Seal the motor holder, handpieces, and contra-angle handpieces (and instru‐
ments if applicable) in a sterilisation bag and sterilise them.
Note
Allow the sterilised items to cool to room temperature before using them again.
2.5
Reprocessing of the assistant unit
2.5.1
Cleaning and disinfection of assistant element
▶ Remove the instruments and suction hoses from the holder.
Care instructions KaVo Primus® 1058 S/TM/C