CAUTION
Damage caused by thermodisinfection.
▶ Do not subject the tumbler holder to thermodisinfection.
The tumbler filler can be subjected to thermodisinfection.
Sterilisation of the tumbler holder and filler
CAUTION
Damage caused by sterilisation.
▶ Do not sterilise the tumbler holder and filler.
2.6.3
Preparation of the spittoon
CAUTION
Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
▶ Make sure that the unit is turned off before removing the tumbler filler or the
spittoon bowl.
▶ Always check the correct seating of tumbler filler and spittoon bowl before
turning on the unit.
Cleaning and disinfection of the spittoon bowl
CAUTION
Damage caused by inappropriate cleaning agents.
▶ Do not use acid-containing, sand-containing, and form-forming cleaning
agents.
▶ Add a dose of DEKASEPTOL gel to the spittoon bowl and distribute the agent
with a brush.
▶ For cleaning, allow the gel to act for approx. 5 minutes.
▶ For disinfection, allow the gel to act for approx. 15 minutes.
▶ After letting it work, press the "Spittoon bowl" button.
▶ Clean the spittoon bowl thoroughly using a brush.
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit
20/35