Sterilize the instrument holder
CAUTION
Damage caused by sterilisation.
▶ Do not sterilise the instrument holder.
2.4.4
Setting up the triple-function and multifunctional handpiece
Note
Wet the O-rings of the sleeves after sterilisation using silicone grease
(Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510). After replac‐
ing the cannula, the air channel must be blown out before the start of treatment
to ensure that the air exiting during treatment is dry.
Cleaning the triple-function and multifunction handpiece
Automatic cleaning
Not applicable.
CAUTION
Damage caused by improper cleaning/disinfection.
▶ Do not wash the gripping sleeve and the cannula in the thermodisinfector.
▶ Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶ Do not place the gripping sleeve and the cannula in a disinfectant or ultra‐
sonic bath.
Manual cleaning
Note
Combine manual cleaning with disinfection.
Cleaning the interior
▶ Leave the gripping sleeve and cannula on the instrument, and activate the air
and water for 60 seconds.
External cleaning
CAUTION
Danger of injury if the valve body is hot (heating elements, high-pressure lamp).
Risk of burn injury.
▶ Switch off the unit at the main switch.
▶ Let the handpiece cool after long use.
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
11/35