Step
Description
7.0
Germ reduction of the triple-function and multifunctional handpiece and
the spittoon bowl
Duration: 60 seconds
6.0
Disinfection of the handpieces
Duration: 30 seconds each
5.0
Disinfection of the tumbler
Duration: 40 seconds
4.0
Dwell time (acoustic signal, switch off is possible)
Duration: 30 minutes
3.0
Rinsing of the tumbler
Duration: 40 seconds
2.0
Rinsing of the handpieces
Duration: 30 seconds
1.0
Rinsing of the triple-function and multifunctional handpiece
Duration: 30 seconds
0.1
Waiting position for setting down the handpieces
After intensive germ reduction is over, the "Device ON" LED and the "Service"
LED flash.
▶ Put all handpieces in holders and take off the rinsing attachment.
Interrupt intensive germ reduction during the dwell time
So that no one will have to remain with the device throughout the entire intensive
germ reduction process, intensive germ reduction can be interrupted in program
step 4 (e.g. on the last workday before the weekend). Program step 4 is the dwell
time, in which the microorganisms are killed by an increased OXYGENAL 6 con‐
centration.
Requirement
intensive germ reduction is in program step 4.
You hear a beep for 4 seconds.
▶ Turn the device off.
When the device is switched on after 30 minutes of dwell time, the remaining pro‐
gram steps run immediately and automatically.
If the device is turned on before the 30 minute period is elapsed, the dwell times
finishes first before the remaining steps of the intensive germ reduction program
run.
Skip exposure phase
This function is required when the treatment unit needs to be used (for example in
an emergency, at hospitals, etc.) even though the germ reduction procedure has
already started.
Requirement
The intensive germ reduction is in step 7 to 5 prior to the reaction phase.