Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The KaVo NOMAD PRO 2 Intraoral X-ray System is intended for use in the electromagnetic environment
specified below. The customer or user of the KaVo NOMAD PRO 2 Intraoral X-ray System should ensure that
it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment
– Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
6Vrms in ISM bands
between 0,15 MHz
and 80 MHz
3 V/m
80 MHz to 2,7 GHz
(V1)=3Vrms
(E1)= 6Vrms in ISM
bands
(E1)=3V/m
Portable and mobile communications
equipment should be separated from
the KaVo NOMAD PRO 2 Intraoral
X-ray System by no less than the
distances calculated/listed below:
D=(3.5/V1)(Sqrt P)
150kHz to 80MHz
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
where P is the max power in watts
and D is the recommended separation
distance in meters.
Field strengths from fixed transmitters,
as determined by an electromagnetic
site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the vicinity of
equipment containing a transmitter.
Where P is the maximum output
power rating of the
transmitter
in watts (W) according to the
transmitter manufacturer
and D is
the recommended separation distance
in meters (m).