31
30
A
lerts
& A
lArms
: V
isuAl
And
A
udible
i
ndicAtors
F
unction
/r
esolution
V
isuAl
A
udible
Low Battery
Alarm
5 audible tones
The
Recharge Handset
alarm
terminates after five seconds
and goes into auto shutdown.
Replace the current handset with
a freshly charged handset.
NOTE:
If the battery voltage is lower than
required for the X-ray exposure, the
device will not allow the exposure.
Incomplete
Exposure
Alarm
A warning tone
This alarm activates if the trigger
is released before the timed X-ray
exposure finishes. This condition is
cleared by pressing any button on
the user interface panel, by turning
the power off then on. To avoid this
error, depress the trigger for the
duration of the exposure time.
System Failure
Alarm
Descending tone
See
Section 6.2 Troubleshooting
(6.2.8) or an KaVo authorized
service center.
Handset
Failure Alarm
5 audible tones
See
Section 6.2 Troubleshooting
(6.2.12) or an KaVo authorized
service center.
The KaVo NOMAD PRO 2 should not be operated if it has been dropped or if
performance degrades; it should be returned to KaVo for an evaluation.