6.0 KaVo NOMAD PRO 2 Care and Upkeep
6.1 Alarms and Alerts
The visual and audible
alarms
signal a programmed action designed to prevent harm to
operators, patients, and/or the KaVo NOMAD PRO 2.
The visual/audible
alerts
confirm
normal conditions or draw the operator’s attention to a required action.
NOTE:
All audible signals except X-ray termination and completion may be turned down
or off in the
Audible Sounds
menu. See
Section 4.10 Settings Menu
.
A
lerts
& A
lArms
: V
isuAl
And
A
udible
i
ndicAtors
F
unction
/r
esolution
V
isuAl
A
udible
Enabling Alert
None
This alert is initiated by a single pull
and release of the trigger, within
one second. If no further action
is taken, the device will shift into
the
Ready
state. If the trigger is
pulled again before the
Ready
state is indicated, the second pull is
disregarded.
Ready Alert
Double ascending
tone
At the conclusion of the
Enabling
X-Rays
alert, the device will shift
into the
Ready
state. The
Ready
state will last for 15 seconds, or
until one of the buttons on the user
interface panel is pressed, or until
the exposure is initiated by a second
pull of the trigger.