Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The KaVo NOMAD PRO 2 Intraoral X-ray System is intended for use in the electromagnetic environment
specified below. The customer or user of the KaVo NOMAD PRO 2 Intraoral X-ray System should ensure that
it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF emissions
CISPR 11
Group 1
The
KaVo NOMAD PRO 2
Intraoral X-ray System uses RF energy
only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The
KaVo NOMAD PRO 2
Intraoral X-ray System is suitable for
use in all establishments, including domestic, and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies