9
Data on electromagnetic compatibility according to EN60601-1-2
9.1
Electromagnetic Transmissions
The Globus 1078 S treatment unit is for use in an environment like the one cited
below. The customer or user of the Globus 1078 S should ensure that it is used in
the correct environment.
Measurements of noise transmis‐
sions
Conformance
Electromagnetic environment -
Guidelines
HF transmission according to
CISPR 11
Group 1
The Globus 1078 S uses HF energy
only for its internal operation. Its HF
transmission is therefore very low,
and it is improbable that neighbour‐
ing electronic devices will be distur‐
bed.
HF transmission according to
CISPR 11
Class B
The Globus 1078 S is for use in all
facilities including residential ones,
and facilities that are directly con‐
nected to a public power supply that
also supplies residential buildings.
Transmissions of harmonics ac‐
cording to IEC 61000-3-2
Class A
The Globus 1078 S is for use in all
facilities including residential ones,
and facilities that are directly con‐
nected to a public power supply that
also supplies residential buildings.
Transmission of voltage fluctuations
or flicker according to IEC
61000-3-3
Conforms
The Globus 1078 S is for use in all
facilities including residential ones,
and facilities that are directly con‐
nected to a public power supply that
also supplies residential buildings.
9.2
Resistance to electromagnetic interference
The Globus 1078 S treatment unit is for use in an environment like the one cited
below. The customer or user of the Globus 1078 S should ensure that it is used in
the correct environment.
Immunity tests
IEC 60601 test level
Conformance level
Electromagnetic environ‐
ment - guidelines
Electrostatic discharge
(ESD) according to IEC
61000-4-2
± 6 kV contact discharge
± 8 kV atmospheric dis‐
charge
± 2/4/6 kV contact dis‐
charge
± 2/4/8 kV atmospheric
discharge
Floors should be made of
wood or concrete or have
ceramic tiles. When the
floor is made of synthetic
material, the relative hu‐
midity must be at least
30%.
Fast transient electrical
disturbances/ Bursts ac‐
cording to IEC 61000-4-4
± 2 kV for power lines
± 1 kV for input and output
lines
± 2 kV for power lines
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
Instructions for use Globus 1078 S
9 Data on electromagnetic compatibility according to EN60601-1-2 | 9.1 Electromagnetic Transmissions
73/75