The chair positions can be saved on two footswitches; the standard setting is as
follows:
▪ "Spray default" footswitch: "LP" automatic position (last position)
▪ "Blown air" footswitch: "SP" automatic position (rinsing position)
The footswitch assignment can be changed.
▶ Press the "AP 0", " AP 1", " AP 2" or "SP" buttons on the control element until
you hear a beep.
▶ Simultaneously press the foot pedal and the "Preselected spray" or "Blown air"
foot switch.
The selected automatic position is saved on the footswitch.
Retrieve the saved chair position
The saved positions of the chair (so-called automatic positions) can be retrieved
with the press of a button. Four automatic positions can be retrieved with the controls
on the dentist's unit and assistant's unit, and two automatic positions can be re‐
trieved with the foot control.
See also: Save the chair position, Page 40
CAUTION
Danger of crushing during automatic chair movement.
The patient or treatment personnel can be clamped.
▶ Monitor the patient and treatment personnel when changing the chair position.
Retrieve the chair positions using the control element
▶ Briefly press the LP/AP button.
The LEDs of the buttons "AP 0", " AP 1", " AP 2" and "SP" flash for about four
seconds.
▶ During these four seconds, briefly press the "AP 0", " AP 1", " AP 2" or "SP"
buttons.
The chair moves into the automatic position.
Instructions for use Globus 1078 S
4 Operation | 4.3 Adjusting the patient chair
41/75