Removing the cannulas
▶ Remove the cannula with a slight amount of rotation while holding the handpiece
on the grip sleeve.
Setting the cold-light intensity
The cold light intensity is adjusted in the same manner as with the turbine.
Changing the high-pressure lamp
CAUTION
Injury hazard from hot valve body.
Burns from contact.
▶ Switch main device switch off.
▶ Let the lamp cool down after long use.
▶ Pull off the grip sleeve ④ together with the cannula from the valve body ①.
▶ Push the holder ③ forward, and pull the defective high-pressure bulb ② out of
the socket.
▶ Insert the new high-pressure lamp (Mat. no. 1.002.2928).
Check the position of the contacts.
▶ Pull back the holder ③ slightly.
▶ Push the grip sleeve and cannula on together until they audibly lock into place.
Instructions for use Globus 1078 S
4 Operation | 4.8 Using instruments
60/75