Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
Accessories required:
▪ Tap water 30
o
C ± 5
o
C
▪ Brush, e.g. medium-hard toothbrush
▶
Disassemble the gripping sleeve and cannula.
▶
Detach the key part from the gripping sleeve.
▶
Clean the key part, gripping sleeve and cannula by brushing them under
running water (at least tap water quality).
▶
Immediately remove contamination from materials used in the dental prac-
tice (impression materials, caustic chemicals).
▶
Carefully wipe off contamination from the cannula tip with a soft Q-tip or
soft cloth and water.
NOTICE
Damage caused by using a nozzle needle that is too long.
Damage to the internal hoses of the cannula.
▶
Use a short nozzle needle (
Mat. no. 1.004.4986
) to clean the media exit
holes.
▶
Carefully remove contamination from the media exit ports using the short
nozzle cleaning needle.
Disinfection of the triple-function and multifunction
handpiece
Manual disinfection
External disinfection
▶
Wipe-disinfect the surfaces with a soft cloth.
Automatic disinfecting
Not applicable.
NOTICE
Damage caused by improper cleaning/disinfection.
▶
Do not wash the gripping sleeve and the cannula in the washer disinfector.
▶
Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶
Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.