Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
14 / 48
▶
Wipe the instrument arm S table with disinfectant.
Sterilisation of instrument arm S table
NOTICE
Damage caused by sterilisation.
▶
Do not sterilise the instrument arm S table.
2.4.4 Setting up the triple-function and multifunction
handpiece
Note
Wet the O-rings of the gripping sleeves after sterilisation using silicone
grease (
Mat. no. 1.000.6403
) or KaVo Rota Spray 2 (
Mat. no.
0.411.7510
). After replacing the cannula, the air channel must be blown out
before the start of treatment to ensure that the air exiting during treatment
is dry.
Cleaning the triple-function and multifunction handpiece
Automatic cleaning
Not applicable.
NOTICE
Damage caused by improper cleaning/disinfection.
▶
Do not wash the gripping sleeve and the cannula in the washer disinfector.
▶
Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶
Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.
Manual cleaning
Note
Combine manual cleaning with disinfection.
Cleaning the interior
▶
Leave the gripping sleeve and cannula on the instrument, and activate the
air and water for 60 seconds.
External cleaning
CAUTION
Danger of injury if the valve body is hot (heating elements, high-pres-
sure lamp).
Risk of burn injury.
▶
Switch off the unit at the main switch.
▶
Let the handpiece cool after long use.