Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
20 / 48
▪ Motor and motor cable
Note
Please comply with the corresponding Instructions for use when you re-pro-
cess the motor and motor cable.
The re-processing steps for the handpiece tray are as follows:
KaVo recommends washer disinfectors in compliance with ISO 15883-1, which
are operated using alkaline cleaning agents.
▶
For program settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the thermodisinfector.
Drying
Note
Coolant hose with accessories is intended for single use only and is not to be
disinfected and sterilised. No drying required.
▶
Allow all disinfected and and sterilised parts to dry fully on room air before
using them again.
Automatic Drying
The drying procedure is normally part of the cleaning program of the thermod-
isinfector.
▶
Please observe the instructions for use of the thermodisinfector.
Service, inspection and testing after preparation
Note
It is essential to comply with the hygiene requirements (sterility) during the
test after reprocessing. If sites of fracture and clear changes of the surface
are visible, the parts need to be checked by the Service.
Check for cleanliness, intactness, servicing, and repair as described in the fol-
lowing:
▶
Check the adjustable functions of the unit and the motor function.
▶
Check the hose pump for sufficient coolant flow rate.
▶
Check the control commands on the wireless foot control.
Packaging the assembly kit for physiological saline
solution
Note
The quality and use of the sterilisation packaging must comply with applic-
able standards and be suitable for the sterilisation procedure!
Note
If potentially infectious liquids and particles can contact the products, it is re-
commendable to cover and protect these areas with sterile disposable
products.
▶
Seal the handpiece tray and motor cable in a sterilisation pouch.