Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.5 Preparation of the assistant unit
Cleaning and disinfection of the suction hose connector
▶
Clean the disassembled parts under running water.
▶
Thermodisinfect the disassembled parts for at least 10 minutes at 93 °C
(total cycle: approximately 20 minutes)
Sterilising the suction hose connector
NOTICE
Damage caused by sterilisation.
▶
Do not sterilise the suction hose connector.
Note
Sterilise the suction cannulas according to the manufacturer's instructions.
2.5.3 Reprocessing of sieve inserts and suction hoses
Note
Check the sieve inserts daily and after each surgery. The sieve inserts must
be replaced at the latest when the suction decreases (
Mat. no.
0.763.3423
).
Cleaning is not recommended for hygienic reasons.
▶
Press the top edge of the service flap on the device base to open it.
▶
Hold the napkin under the sieve seat, and pull the sieve insert out of the
housing.
25 / 48