53
1.
Stand on a small step to look directly down each of the chimneys. Their
inside wall should be smooth and solid white.
2.
If it is deemed necessary to clean the inside of the chimneys, continue to the
next steps.
3.
Turn the instrument off and let it cool down.
4.
Slowly pull each chimney to remove them.
5.
Use a suitable brush or the eraser end of a pencil to scrub any deposit from
the inside wall.
6.
Carefully re-install the chimneys after cleaning.
Element
terminal
connections
IMPORTANT
: Always disconnect the instrument from the power outlet when
working with panels removed. 240 Volts inside! Risk of electrocution!
1.
To check the integrity of the element ceramic connectors and terminals, do
the following:
1.1
Unplug instrument and let the furnace cool down completely.
1.2
Remove the element connection access panel (refer to page 60,
for details on this step).
1.3
Visually inspect and verify that terminals of all ceramic connectors are
not overly oxidized. If a faulty connector is found, replace this ceramic
connector immediately.
1.4
Using an appropriate tool, verify that all wire terminals of ceramic
connectors are well tightened, both on the element side and the wire
side. Do not over-torque the screws. Simply verify that they have not
become loosened.
Furnace
cleanliness
IMPORTANT
: Always disconnect the instrument from the power outlet when
working with panels removed. 240 Volts inside! Risk of electrocution!
1.
To run this inspection, instrument must be turned OFF, unplugged and furnace
at room temperature.
1.1
Unplug the X-300 and let it cool down completely.
1.2
Remove the instrument’s main cover (see page 62,
removal for details).
1.3
The furnace door should be hanging ajar.